National Instruments NI PXI-5422 Manuel d'utilisateur Page 56

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 63
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 55
NI PXI-5422 Calibration Procedure 56 ni.com
5. Call niFgen_SetAttributeViInt32 to set the offset DAC value
(niFgen property node: Calibration»Offset DAC Value) using the
following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_OFFSET_DAC_VALUE
value: 32767
6. Call niFgen_SetAttributeViBoolean to set the analog filter state
(niFgen property node: Output Attributes»Analog Filter Enabled)
using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_ANALOG_FILTER_ENABLED
value: VI_FALSE
7. Call niFgen_SetAttributeViReal64 to set the pre-amplifier
attenuation (niFgen property node: Calibration»Pre-Amplifier
Attenuation) using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_PRE_AMPLIFIER_ATTENUATION
value: 0
8. Call niFgen_SetAttributeViReal64 to set the post-amplifier
attenuation (niFgen property node: Calibration»Post-Amplifier
Attenuation) using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
value: 0
9. Call niFgen_SetAttributeViReal64 to set the output impedance
(niFgen property node: Basic Operation»Output Impedance) using
the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
channelName: "0"
attributeID: NIFGEN_ATTR_OUTPUT_IMPEDANCE
value: 50
10. Call niFgen_SetAttributeViBoolean to enable the analog output
(niFgen property node: Basic Operation»Output Enabled) using
the following parameters:
Vue de la page 55